Computed tomography dose index

From testwiki
Revision as of 15:01, 22 December 2019 by imported>Citation bot (Alter: last2, pages. Add: pmid, bibcode. Removed URL that duplicated unique identifier. Removed accessdate with no specified URL. Removed parameters. Formatted dashes. | You can use this bot yourself. Report bugs here. | Activated by User:Grimes2 | via #UCB_webform)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

The computed tomography dose index (CTDI) is a commonly used radiation exposure index in X-ray computed tomography (CT), first defined in 1981.[1][2] The unit of CTDI is the gray (Gy) and it can be used in conjunction with patient size to estimate the absorbed dose. The CTDI and absorbed dose may differ by more than a factor of two for small patients such as children.[3]

Definitions

Because CT scanners typically acquire multiple slices during a single rotation with a single beam, the CTDI is calculated by integrating over the dose profile for a single axial rotation, then dividing by the nominal beam width:[4]

CTDI=1nTz+zD(z)dz

where n is the number of slices acquired per single axial rotation, T is the width of a single acquired slice (and thus nT is the nominal beam width) and D(z) is the radiation dose measured at position z along the scanner's main axis - the dose profile.

This measurement is most often made using a 100-mm standard pencil dose chamber as this is representative of a typical scan length:

CTDI100=1nT50mm50mmD(z)dz.

The absorbed dose to water Dw(z) (used to refer back to patient dose) is typically measured in a cylindrical head (16 cm diameter) or body (32 cm diameter) phantom of length approximately 14–15 cm.[2]

The dose distribution imparted by a CT scan is much more homogeneous than that imparted by radiography, but is still somewhat larger near the skin than in the centre of the body. The weighted CTDI was introduced to account for this:[5]

CTDIw=13CTDI100central+23CTDI100peripheral

using measurements acquired at central and peripheral positions in the head or body phantoms described above.

CTDI in helical CT

In helical CT, the pitch of the machine - a factor of the speed at which the couch travels through the gantry and the tube rotation frequency - also impacts on patient dose. The pitch factor, P, is defined as[6]

P=d360C

where d360 is the distance travelled by the couch during one full gantry rotation and C is the beam collimation (single-slice CT) or the total thickness of all simultaneously acquired slices (multislice CT). The following quantity is therefore used to take account of pitch:

CTDIvol=CTDIwP

Similar measures with yet wider chambers are useful for CT systems with large numbers of detector rows.[7]

CTDI can also be measured with polymer gel dosimetry.[8]

Relation to DLP

The dose-length product (DLP) is a quantity defined for use in CT as

DLP=CTDIvol.nT

for n and T as described above (nT is therefore the total scan length). This quantity is analogous to the dose-area product (DAP) used in planar radiography.

References